Skip to main content
NABL Accredited ISO/IEC 17025:2017 • CC-4178 Permanent Lab Calibration 30 Scopes

Volume Calibration Services

Burettes, Pycnometers, Separating Funnels, Measuring Cylinders, Graduated Jars, Density Bottles, Centrifuge Tubes, Micropipettes, Piston Pipettes, Single/Graduated Pipettes, Dispensers, Volumetric Flasks, Le Chatelier Flasks, Standard Flasks, Conical Flasks, Beakers & Specific Gravity Bottles.

WhatsApp Lab Desk
Calibration Capability (CMC)
Expanded Uncertainty: ± 0.025 µl to ± 5.8 ml (0.1 µl to 5000 ml).
Primary Master Equipment
Electronic Micro Balance (Readability: 0.001 mg), Electronic Semi Micro Balance (Readability: 0.01 mg), Electronic Precision Weighing Balances (Readability: 1 mg & 100 mg), Degassed ASTM Type 1 Deionized Water System.
Governing Standards
ISO 4787:2021, ISO 8655-6:2022, ISO 3507:1999.
Key Parameters & Applications
Gravimetric dispensed and containment volume, deliverable capacity, systematic error (inaccuracy), random error as per ISO 4787:2021, ISO 8655-6:2022 & ISO 3507:1999.

Instruments & Parameters Covered (30)

Search or review approved measurement ranges and Calibration & Measurement Capabilities (CMC).

Instrument / Parameter Calibration Range CMC (±) Master / Method
Burette, Pycnometer, Separating Funnel (> 10 ml to 50 ml) > 10 ml to 50 ml ± 7.4 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021 & ISO 3507:1999
Burette, Pycnometer, Separating Funnel (> 50 ml to 100 ml) > 50 ml to 100 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021 & ISO 3507:1999
Burette, Pycnometer, Separating Funnel (0.1 ml to 10 ml) 0.1 ml to 10 ml ± 7.4 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021 & ISO 3507:1999
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (> 10 ml to 50 ml) > 10 ml to 50 ml ± 7.4 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021 & ISO 3507:1999
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (> 100 ml to 1000 ml) > 100 ml to 1000 ml ± 0.05 ml Using Electronic Weighing Balance (Readability: 1 mg) by Gravimetric Method as per ISO 4787:2021
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (> 1000 ml to 2000 ml) > 1000 ml to 2000 ml ± 0.05 ml Using Electronic Weighing Balance (Readability: 1 mg) by Gravimetric Method as per ISO 4787:2021
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (> 2000 ml to 4000 ml) > 2000 ml to 4000 ml ± 5.8 ml Using Electronic Weighing Balance (Readability: 1 mg) by Gravimetric Method as per ISO 4787:2021
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (> 50 ml to 100 ml) > 50 ml to 100 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Measuring Cylinder, Measuring Jar, Graduated Jar, Density Bottle, Crow Receiver, Centrifuge/Filter Tube (1 ml to 10 ml) 1 ml to 10 ml ± 2 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021 & ISO 3507:1999
Micropipette, Piston Pipette (> 100 µl to 1000 µl) > 100 µl to 1000 µl ± 4.2 µl Using Electronic Micro Balance (Readability: 0.001 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (> 5000 µl to 10000 µl) > 5000 µl to 10000 µl ± 6 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (> 10 µl to 100 µl) > 10 µl to 100 µl ± 0.52 µl Using Electronic Micro Balance (Readability: 0.001 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (> 1000 µl to 5000 µl) > 1000 µl to 5000 µl ± 4.2 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (0.1 µl to 0.5 µl) 0.1 µl to 0.5 µl ± 0.025 µl Using Electronic Micro Balance (Readability: 0.001 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (0.5 µl to 1.0 µl) 0.5 µl to 1.0 µl ± 0.025 µl Using Electronic Micro Balance (Readability: 0.001 mg) by Gravimetric Method as per ISO 8655-6:2022
Micropipette, Piston Pipette (1 µl to 10 µl) 1 µl to 10 µl ± 0.03 µl Using Electronic Micro Balance (Readability: 0.001 mg) by Gravimetric Method as per ISO 8655-6:2022
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (> 25 ml to 50 ml) > 25 ml to 50 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (> 50 ml to 100 ml) > 50 ml to 100 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (0.01 ml to 0.1 ml) 0.01 ml to 0.1 ml ± 1.7 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (0.1 ml to 1 ml) 0.1 ml to 1 ml ± 1.7 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (> 1 ml to 10 ml) > 1 ml to 10 ml ± 7.4 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (> 10 ml to 25 ml) > 10 ml to 25 ml ± 7.4 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Pipette, Single Volume Pipette, Graduated Volume Pipette, Dilution Pipette, Dispenser (> 100 ml to 1000 ml) > 100 ml to 1000 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Volumetric Flask, Le Chatelier Flask, Standard Flask, Conical Flask, Beaker, Specific Gravity Bottle, Dispenser (> 10 ml to 50 ml) > 10 ml to 50 ml ± 10.1 µl Using Electronic Semi Micro Balance (Readability: 0.01 mg) by Gravimetric Method as per ISO 4787:2021
Volumetric Flask, Le Chatelier Flask, Standard Flask, Conical Flask, Beaker, Specific Gravity Bottle (> 200 ml to 500 ml) > 200 ml to 500 ml ± 0.04 ml Using Electronic Weighing Balance (Readability: 1 mg) by Gravimetric Method as per ISO 4787:2021

Frequently Asked Questions

Are Volume calibrations NABL accredited?

Yes. Our Volume laboratory is accredited under ISO/IEC 17025:2017 by NABL (Certificate No: CC-4178), ensuring global acceptance and unbroken traceability to NPL/NIST primary standards.

Do you offer on-site Volume calibration at client plants?

Calibration for Volume is performed inside our environmentally controlled permanent metrology facility in Madipakkam, Chennai, to preserve measurement stability and CMC accuracy.

What is the standard turnaround time for calibration certificates?

Standard turnaround time is 3 to 5 working days for permanent laboratory orders. Express same-day and next-day calibration slots are available upon prior request.

How do I request a quotation for Volume?

Click the "Request a Quote" button on this page to pre-select Volume, upload your instrument list (up to 15 MB PDF/Excel), or call +91 98406 72352 for immediate scope verification.

Explore Related Calibration Disciplines & Services

Industrial Turnkey AMC Plans • NABL Lab Setup Consultancy • ISO/IEC 17025:2017 Training Courses • Direct Lab Helpdesk
Call Lab WhatsApp